Language Dictionary - 4Command Syntax [SOURce:]VOLTage:RANGe Parameters 0 through MAX | MINimum | MAXimumUnit V (volts)*RST Value MAXimum (high range)Examples VOLT:RANG 15 SOUR:VOLT:RANGE MINQuery Syntax [SOURce:]VOLTage:RANGe?Returned Parameters Related Commands VOLT VOLT:SLEW[SOURce:]VOLTage:SLEWChannel SpecificThis command sets the slew rate for all programmed changes in the input voltage level of the electronicload. This command programs both positive and negative going slew rates. Although any slew rate valuemay be entered, the electronic load selects a slew rate that is closest to the programmed value.MAXimum sets the slew to the fastest possible rate. MINimum sets the slew to the slowest rate. Slewrates less than the minimum value are set to MINimum. Slew rates greater than the maximum value areset to MAXimum.Command Syntax [SOURce:]VOLTage:SLEW[:BOTH] Parameters 0 to 9.9E37 | MAXimum | MINimumUnit V (volts per second)*RST Value MAXimumExamples VOLT:SLEW 50 VOLT:SLEW MAXQuery Syntax [SOURce:]VOLTage:SLEW[:BOTH]?Returned Parameters Related Commands VOLT:SLEW:POS VOLT:SLEW:NEG[SOURce:]VOLTage:SLEW:NEGativeChannel SpecificThis command sets the slew rate of the voltage for negative going transitions. MAXimum sets the slew tothe fastest possible rate. MINimum sets the slew to the slowest rate.Command Syntax [SOURce:]VOLTage:SLEW:NEGative Parameters 0 to 9.9E37 | MAXimum | MINimumUnit V (volts per second)*RST Value MAXimumExamples VOLT:SLEW:NEG 50 VOLT:SLEW:NEG MAXQuery Syntax [SOURce:]VOLTage:SLEW:NEGative?Returned Parameters Related Commands VOLT:SLEW69