4 - Language Dictionary[SOURce:]VOLTage:SLEW:POSitiveChannel SpecificThis command sets the slew rate of the voltage for positive going transitions. MAXimum sets the slew tothe fastest possible rate. MINimum sets the slew to the slowest rate.Command Syntax [SOURce:]VOLTage:SLEW:POSitive Parameters 0 to 9.9E37 | MAXimum | MINimumUnit V (volts per second)*RST Value MAXimumExamples VOLT:SLEW:POS 50 VOLT:SLEW:POS MAXQuery Syntax [SOURce:]VOLTage:SLEW:POSitive?Returned Parameters Related Commands VOLT:SLEW[SOURce:]VOLTage:TLEVelChannel SpecificThis command specifies the transient level of the input voltage. The transient function switches betweenthe immediate setting and the transient level. Refer to Table 4-1 for model-specific programming ranges.Command Syntax [SOURce:]VOLTage:TLEVel Parameters 0 through MAX | MINimum | MAXimumUnit V (volts)*RST Value MAXimumExamples VOLT:TLEV 5 VOLT:TLEV .5Query Syntax [SOURce:]VOLTage:TLEVel?Returned Parameters Related Commands VOLT:[SOURce:]VOLTage:TRIGgeredChannel SpecificThis command sets the voltage level that will become active when the next trigger occurs.Command Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude] Parameters refer to Specifications Table in User’s GuideUnit V (volts)*RST Value MAXimumExamples VOLT:TRIG 120 VOLT:LEV:TRIG 150Query Syntax [SOURce:]VOLTage[:LEVel]:TRIGgered[:AMPLitude]?Returned Parameters (if the trigger level is not programmed, the immediatelevel is returned)Related Commands VOLT VOLT:MODE70