Language Dictionary - 4SENSe:CURRent:RANGeChannel SpecificThis command sets the current measurement range. There are two current measurement ranges:High Range: model dependent, see Table 4-1Low Range: model dependent, see Table 4-1A value of infinity is returned if the measured value is outside the specified current measurement range.Command Syntax SENSe:CURRent:RANGe Parameters 0 through MAX | MINimum | MAXimumUnit A (amperes)*RST Value MAX (high range)Examples SENS:CURR:RANGE MINQuery Syntax SENSe:CURRent:RANGe?Returned Parameters SENSe:SWEep:POINtsChannel SpecificThis command specifies how many data points are taken in any measurement. Applies to both voltageand current measurements. The number of points can be specified, from 1 to 4096.Command Syntax SENSe:SWEep:POINts Parameters 1 through 4096 | MINimum | MAXimum*RST Value 1000Examples SENS:SWE:POIN 2048Query Syntax SENSe:SWEep:POINts?Returned Parameters Related Commands SENS:SWE:TINT MEAS:ARRSENSe:SWEep:OFFSetChannel SpecificThis command specifies a delay time after the trigger, but before the measurement is taken. The delay isspecified in seconds. The measurement offset can be either positive or negative with respect to thetrigger.NOTE: Negative measurement offsets can only be programmed in conjunction with delayedtriggers. The negative measurement offset cannot exceed the trigger delay value.Command Syntax SENSe:SWEep:OFFSet Parameters 0 through 0.032 | MINimum | MAXimumUnit seconds*RST Value 0 (zero)Examples SENS:SWE:OFFS 0.01Query Syntax SENSe:SWEep:OFFSet?Returned Parameters Related Commands SENS:SWE:TINT MEAS:ARR75