4 - Language Dictionary[SOURce:]LIST:RESistance:TLEVel[SOURce:]LIST:RESistance:TLEVel:POINTs?Channel SpecificThis command specifies the transient resistance level for each step. The transient function switchesbetween the immediate setting and the transient level. LIST:RESistance:TLEVel:POINts? returns thenumber of points programmed.Command Syntax [SOURce:]LIST:RESistance:TLEVel {,}Parameters refer to Table 4-1Unit Ω (ohms)*RST Value MAXimumExamples LIST:RES:TLEV 5 LIST:RES:TLEV .5Query Syntax [SOURce:]LIST:RESistance:TLEVel?[SOURce:]LIST:RESistance:TLEVel:POINts?Returned Parameters {,}Related Commands RES:TLEV[SOURce:]LIST:STEPThis command specifies how the list sequencing responds to triggers. The following parameters may bespecified. This command is not channel specific, it applies to the entire mainframe.ONCE Causes the list to advance only one point after each trigger. Triggers that arrive during adwell delay are ignoredAUTO Causes the entire list to be executed sequentially after the starting trigger, paced by itsdwell delays. As each dwell delay elapses, the next point is immediately executed.Command Syntax [SOURce:]LIST:STEP Parameters ONCE | AUTO*RST Value AUTOExamples LIST:STEP ONCEQuery Syntax [SOURce:]LIST:STEP?Returned Parameters Related Commands LIST:COUN LIST:DWEL[SOURce:]LIST:TRANsient[SOURce:]LIST:TRANsient:POINts?Channel SpecificThis command turns the transient generator on or off for each step. LIST:TRANsient:POINts? returns thenumber of points programmed.Command Syntax [SOURce:]LIST:TRANsient[:STATe] {,}Parameters 0 | 1 | OFF | ON*RST Value OFFExamples LIST:TRAN 1 LIST:TRAN 0Query Syntax [SOURce:]LIST:TRANsient[:STATe]?[SOURce:]LIST:TRANsient:POINts?Returned Parameters {,}Related Commands TRAN84