Language Dictionary - 4[SOURce:]CURRent:SLEWChannel SpecificThis command sets the slew rate for all programmed changes in the input current level of the electronicload. This command programs both positive and negative going slew rates. Although any slew rate valuemay be entered, the electronic load selects a slew rate that is closest to the programmed value.MAXimum sets the slew to the fastest possible rate. MINimum sets the slew to the slowest rate. Slewrates less than the minimum value are set to MINimum. Slew rates greater than the maximum value areset to MAXimum.Command Syntax [SOURce:]CURRent:SLEW[:BOTH] Parameters 0 to 9.9E37 | MAXimum | MINimumUnit A (amps per second)*RST Value MAXimumExamples CURR:SLEW 50 CURR:SLEW MAXQuery Syntax [SOURce:]CURRent:SLEW[:BOTH]?Returned Parameters Related Commands CURR:SLEW:NEG CURR:SLEW:POS[SOURce:]CURRent:SLEW:NEGativeChannel SpecificThis command sets the slew rate of the current for negative going transitions. MAXimum sets the slew tothe fastest possible rate. MINimum sets the slew to the slowest rate.Command Syntax [SOURce:]CURRent:SLEW:NEGative Parameters 0 to 9.9E37 | MAXimum | MINimumUnit A (amps per second)*RST Value MAXimumExamples CURR:SLEW:NEG 50 CURR:SLEW:NEG MAXQuery Syntax [SOURce:]CURRent:SLEW:NEGative?Returned Parameters Related Commands CURR:SLEW[SOURce:]CURRent:SLEW:POSitiveChannel SpecificThis command sets the slew rate of the current for positive going transitions. MAXimum sets the slew tothe fastest possible rate. MINimum sets the slew to the slowest rate.Command Syntax [SOURce:]CURRent:SLEW:POSitive Parameters 0 to 9.9E37 | MAXimum | MINimumUnit A (amps per second)*RST Value MAXimumExamples CURR:SLEW:POS 50 CURR:SLEW:POS MAXQuery Syntax [SOURce:]CURRent:SLEW:POSitive?Returned Parameters Related Commands CURR:SLEW63