Language Dictionary - 4Command Syntax [SOURce:]RESistance:SLEW:NEGative Parameters 0 to 9.9E37 | MAXimum | MINimumUnit Ω (ohms/second)*RST Value MAXimumExamples RES:SLEW:NEG 50 RES:SLEW:NEG MAXQuery Syntax [SOURce:]RESistance:SLEW:NEGative?Returned Parameters Related Commands RES:SLEW[SOURce:]RESistance:SLEW:POSitiveChannel SpecificThis command sets the slew rate of the resistance for positive going transitions. MAXimum sets the slewto the fastest possible rate. MINimum sets the slew to the slowest rate.Command Syntax [SOURce:]RESistance:SLEW:POSitive Parameters 0 to 9.9E37 | MAXimum | MINimumUnit Ω (ohms/second)*RST Value MAXimumExamples RES:SLEW:POS 50 RES:SLEW:POS MAXQuery Syntax [SOURce:]RESistance:SLEW:POSitive?Returned Parameters Related Commands RES:SLEW[SOURce:]RESistance:TLEVelChannel SpecificThis command specifies the transient level of the resistance. The transient function switches between theimmediate setting and the transient level. Refer to Table 4-1 for model-specific programming ranges.Command Syntax [SOURce:]RESistance:TLEVel Parameters 0 through MAX | MINimum | MAXimumUnit Ω (ohms)*RST Value MAXimumExamples RES:TLEV 5 RES:TLEV .5Query Syntax [SOURce:]RESistance:TLEVel?Returned Parameters Related Commands RES:[SOURce:]RESistance:TRIGgeredChannel SpecificThis command sets the resistance level that will become active when the next trigger occurs.Command Syntax [SOURce:]RESistance[:LEVel]:TRIGgered[:AMPLitude] Parameters refer to Specifications Table in User’s GuideUnit Ω (ohms)*RST Value MAXimumExamples RES:TRIG 120 RES:LEV:TRIG 150Query Syntax [SOURce:]RESistance[:LEVel]:TRIGgered[:AMPLitude]?Returned Parameters (if the trigger level is not programmed, the immediatelevel is returned)Related Commands RES RES:MODE67