4 - Language DictionarySENSe:SWEep:TINTervalChannel SpecificThis command defines the time period between measurement points. The time interval can beprogrammed from 0.00001 to 0.032 seconds in 10 microsecond increments.Command Syntax SENSe:SWEep:TINTerval Parameters 0.00001 - 0.032 | MAXimum | MINimumUnit seconds*RST Value 10 μsExamples SENS:SWE:TINT 100E-6Query Syntax SENSe:SWEep:TINTerval?Returned Parameters Related Commands SENS:SWE:OFFS:POIN MEAS:ARRSENSe:WINDowChannel SpecificThis command sets the window function that is used in dc and ac+dc rms measurement calculations. Thefollowing functions can be selected:HANNing A signal conditioning window that reduces errors in dc and rms measurementcalculations in the presence of periodic signals such as line ripple. It also reducesjitter when measuring successive pulses. The Hanning window multiplies each pointin the measurement sample by the function cosine 4. Do not use the Hanningwindow when measuring single-shot pulses.RECTangular A window that returns measurement calculations without any signal conditioning.NOTE: Neither window function alters the voltage or current data in the measurement array.Command Syntax SENSe:WINDow[:TYPE] Parameters RECTangular | HANNing*RST Value RECTangularExamples SENS:WIND HANNQuery Syntax SENSe:WINDow?Returned Parameters SENSe:VOLTage:RANGeChannel SpecificThis command sets the voltage measurement range. There are two voltage measurement ranges:High Range: model dependent, see Table 4-1Low Range: model dependent, see Table 4-1A value of infinity is returned if the measured value is outside the specified voltage measurement range.Command Syntax SENSe:VOLTage:RANGe[:UPPer] Parameters 0 through MAX | MINimum | MAXimumUnit V (voltage)*RST Value MAX (high range)Examples SENS:VOLT:RANGE MINQuery Syntax SENSe:VOLTage:RANGe?Returned Parameters Related Commands SENS:SWE:TINT MEAS:ARR76