4 - Language Dictionary[SOURce:]LIST:TRANsient:TWIDth[SOURce:]LIST:TRANsient:TWIDth:POINts?Channel SpecificThis command sets the transient pulse width for each step when the generator is in PULSe mode.LIST:TRANsient:TWIDth:POINts? returns the number of points programmed.Command Syntax [SOURce:] LIST:TRANsient:TWIDth {,}Parameters 0.00005s - 4s | MAXimum | MINimumUnit seconds*RST Value 0.0005sExamples LIST:TRAN:TWID .005 LIST:TRAN:TWID 5E-4Query Syntax [SOURce:]LIST:TRANsient:TWIDth?[SOURce:]LIST:TRANsient:TWIDth:POINts?Returned Parameters {,}Related Commands TRAN:TWID[SOURce:]LIST:VOLTage[SOURce:]LIST:VOLTage:POINts?Channel SpecificThis command specifies the voltage setting for each list step. Refer to Table 4-1 for model-specificprogramming ranges. LIST:VOLTage:POINts? returns the number of points programmed.Command Syntax [SOURce:]LIST:VOLTage[:LEVel] {,}Parameters 0 through MAX | MINimum | MAXimumUnit V (volts)Examples LIST:VOLT 2.5,3.0,3.5LIST:VOLT MAX,3.5,2.5,MINQuery Syntax [SOURce:]LIST:VOLTage[:LEVel]?[SOURce:]LIST:VOLTage:POINts?Returned Parameters {,}Related Commands VOLT[SOURce:]LIST:VOLTage:RANGe[SOURce:]LIST:VOLTage:RANGe:POINTs?Channel SpecificThis command sets the voltage range for each list step. There are two voltage ranges.High Range: model dependent, see Table 4-1Low Range: model dependent, see Table 4-1When you program a range value, the load automatically selects the range that corresponds to the valuethat you program. If the value falls in a region where ranges overlap, the load selects the range with thehighest resolution. LIST:VOLTage:RANGe:POINts? returns the number of points programmed.NOTE: If the existing IMMediate, TRANsient, and TRIGgered voltage list settings are outside thenew range, a list error is generated when the list is initiated and the list does not execute.86