STARTUPAxio Scope.A1 Mounting the Standard Components Carl ZeissM60-2-0007 e 05/08 493.1.16 Inserting the Adjustment Aid intothe Upper Stand Part FL/HBO• Remove the cover (447H447HFig. 3-26/1) from themounting aperture of the adjustment aid.• Insert the adjustment aid (4 4 8 H 4 4 8 H Fig. 3-26/4) andtighten the three included screws (449H449H Fig. 3-26/3).• Mount the cap with the opening (450H450HFig. 3-26/2).Make sure it locks.• Slide in the movable connecting piece of theadjustment aid.3.1.17 HBO 50 Illuminator3.1.17.1 Inserting or changing the burnerof the HBO 50 illuminatorBE CAREFUL – Danger of burns!Before changing the burner ofHBO 50 switch off the transformerand pull its plug. Let the lamp casecool off for at least 15 min. After thelamp case has cooled down, removethe HBO 50 from the upper standpart and put it aside on a planesurface.• Loosen the clamping screw (4 5 1 H 4 5 1 H Fig. 3-27/1) withthe ball head screwdriver SW 3.• Press down the release button (452H452HFig. 3-27/2) ofthe HBO 50, pull out the lamp carrier (453H453HFig.3-27/3) and set it aside.Fig. 3-26 Inserting the adjustment aidFig. 3-27 Opening the HBO 50