- 31 / 79 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setDM-CAM with Camera Screw10.3The digital DM-CAM microscope camera is used to monitor the probe tip position and the distancebetween the probe tip and DUT.The camera screw is needed to mount the DM-CAM on the DM-CAM holder.3.Figure 27: Camera screw shown separately and as connected to the DM-CAM holder.3The camera can be fixed by sliding it into the opening of the camera screw and tightening thesmall, silver-coloured knurled screw. The large, silver-coloured knurled screw and the large, black,knurled plastic nut are used to mount the camera screw on the DM-CAM holder.3.Figure 28: DM-CAM microscope camera DM-CAM in the connected stateThe DM-CAM is connected to the HR10 connector marked "Camera" on the scanner's Z-axistower.The DM-CAM's focus can be set by turning its rear knurled nut.