- 61 / 79 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setConnecting the ICR Near-Field Microprobe17.2Important! Please note: The probe tip of the ICR near-field microprobe (ICR probe for short) ishighly sensitive to mechanical stress. Therefore, remove the ICR probe's protective cap only afterthe installation or shortly before the measurement. Do not touch the probe tip. The probe shouldnot be allowed to come into contact with the DUT either!As a prerequisite, the cables must have been laid according to section 17.1.The ICR near-field micro probe has to be installed on the rotary ring of the FLS 106 IC scanner'srotary unit. Remove the knurled screws from the rotary ring.Figure 77: Knurled screws removed from the rotary ringWe recommend that you first connect the two cables to the ICR probe and then mount the probeon the rotary ring.Connect the SMA-SMA measuring cable to the "RF out" connector of the ICR probe (Figure 78)and then the SSMB-SSMB cable to the SSMB connector of the ICR probe (Figure 79).Figure 78: Connecting the SMA cable to the ICR probe Figure 79: Connecting the SSMB cable to the ICR probe