- 66 / 79 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setNow insert the near-field probe into the probe sleeve of the probe holder and carefully tighten thesmall screw on the outside of the probe sleeve by hand (Figure 92).Figure 91: Probe sleeve removed from the magneticholderFigure 92: Near-field probe with probe sleeve in the installedstateThe probe sleeve can also be removed from the holder as it is held magnetically so as to facilitateassembly (section 10.8). Thanks to the rear bore, the probe sleeve can be returned to themagnetic holder of the SH 01 and fastened in the correct position.Finally check that the installed near-field probe moves smoothly in the Z-direction. This ensuresthat the movement is stopped correctly if a collision occurs. Carefully pull the near-field probe upby hand and check its smooth motion for this purpose.Connecting a Near-Field Probe Directly to the FLS 106 IC Scanner18.1.1If operated without a pre-amplifier, the RF measuring cable has to be connected to the near-fieldprobe's measurement output and directly to "RF in" input on the scanner's Z-axis tower (Figure 93).Figure 93: Direct connection of the near-field probe to the scanner