- 63 / 79 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setInstalling the BT 706 Bias-Tee17.3The BT 706 bias-tee has to be connected in the incoming circuit to ensure the ICR probe's powersupply.Mount the bias-tee on the "RF in" input of the scanner's Z-axis tower by hand.Important: Do not use any tools so as to prevent any damage! (Figure 83).Figure 83: BT 706 bias-tee Figure 84: Bias-tee mounted on the scanner ("RF in")Connect the SMA-SMA RA measuring cable that is connected to the SMA coupling to the SMAinput of the BT 706 (Figure 85).Figure 85: Connecting the SMA-SMA cable Figure 86: Connecting the power supply cableConnect the bias-tee power supply cable to the "12 V / DC" input of the bias-tee and to the"12 V DC" connector on the scanner's Z-axis tower (Figure 86).Collision control of the ICR probe17.4The collision control is a protective function in connection with an ICR probe to limit potentialdamage that may be caused if the ICR probe accidentally touches the device under test.The ICR probe may not touch the device under test during operation. The collision control may notbe used to position the ICR probe.