EN-20 RANGER®7000 SCALESTo switch between Over/Under Load, Target WeightTolerance, Target Weight Percentage press the buttoncorresponding to the icon. If desired, edit the valueby using the alphanumerical keypad and press the buttoncorresponding to the icon to save the changes andreturn to the previous screen.Note: The three set limits methods share the same data.4.3.2 Check CountingPress the configuration button and select Check Mode to Check Counting. Place objects on the pan. TheUnder/Accept/Over status is shown in the progress bar area while the actual number of pieces is shown on themain Display Line.The CHECK Home screenMain Display LineReference FieldsFunctionsNote: Refer section 9.5, or press the button for button icon explanation.Set the Average Piece Weight (APW)Note: It is recommended that the APW is larger than 1d. If APWis between 0.05d and 1d, a warning screen will be displayed andthe information line will show ‘Low APW’. If APW is less than0.05d an error screen will appear and the APW value cannot bestored.There are three ways to set the APW, see section 4.2.2 forinstructions.Defining Over/Under LimitsPress the button corresponding to the icon to enterLimit Setup.Note: See section 4.3.1 for information on how to set theOver/Under limits.www. .com information@itm.com1.800.561.8187