RANGER®7000 SCALES EN-697. LEGAL FOR TRADEWhen the scale is used in trade or a legally controlled application it must be set up, verified and sealed inaccordance with local weights and measures regulations. It is the responsibility of the purchaser to ensure that allpertinent legal requirements are met.7.1 SettingsBefore verification and sealing, perform the following steps in order:1. Verify that the menu settings meet the local weights and measures regulations.2. Units menu should be reviewed. Verify the units turned on meet the local weights and measuresregulations.3. Perform a calibration as explained in Section 5.4. Set the position of the Security Switch to the locked position.7.2 VerificationA weights and measures official or authorized service agent must perform the verification procedure.7.3 SealingAfter the scale has been verified, it must be sealed to prevent undetected access to the legally controlled settings.Before sealing the device, ensure that the security switch is in the Locked position.If using a wire seal, pass the sealing wire through the holes in the security screw and tab, as shown.If using a paper seal, place the seal over the flat head screw as shownA. BaseUn-Locked Locked with Wire Seal Locked with Paper SealB. TerminalUn-locked Locked with Wire Seal Locked with Paper SealNote: The Terminal only needs to be sealed if a second scale is attached to the optional 2nd A/D board.String 9: ************************String 10: OHAUS CorporationString 11: 7 Campus Drive Ste 310String 12: Parsippany NJ 07054String 13: www.ohaus.com 1.800.672.7722String 15: User ID:String 16: Project ID:String 17: Scale ID:String 18: Mode:String 19: Signature:____________________String 20: Verified by:_________________Size Acc. % retained5cm 69.96%20mm 92.23%Pan 100.00%Size Acc. Wt. retained5cm 353.7 g20mm 466.3 gPan 505.6 gFineness Modulus: 1.62Signature:____________________Verified by:__________________www. .com information@itm.com1.800.561.8187