EN-50 RANGER®7000 SCALES5.6 Application ModeEnter this sub-menu to enable or disable the desired Scale Applications. Only one application can berunning at a time.Note: The use of each Application is described in detail in Section 4.5.6.1 Turning an Application ON/OFFHighlight the application by pressing the buttonscorresponding to the iconsand then press the button corresponding to the iconto enter the selected submenu.In the Item option screen, enter the Enabled menuto turn it on or off.Once an Application is enabled (turned on) it may be chosen by pressing theApplications button until it’s icon appears in the upper left corner of the home screen.The current menu item status is shown: OFF = disabled, ON = enabled5.7 Weighing UnitsEnter this sub-menu to activate the desired units of measure.Note: Due to national laws, the scale may not include some of the units of measure listed.5.7.1 Units Sub-menuReset kilogram pound gram ounce pound:ounce Custom UnitNote: The setting is locked when the Security Switch is set to the locked position.Custom UnitUse the Custom Unit to display weight in an alternative unit of measure. The custom unit is defined using aconversion factor, where the conversion factor is the number of custom units per gram expressed in scientificnotation (Factor x 10^Exponent).FactorSet the conversion factor using the numeric keypad.Settings of 0.1000000 to 1.9999999 are available. The default setting is 1.0000000.ExponentSet the factor multiplier.-3 = divide the Factor by 1000 (1x10-3)-2 = divide the Factor by 100 (1x10-2)-1 = divide the Factor by 10 (1x10-1)0 = multiply the Factor by 1 (1x100)1 = multiply the Factor by 10 (1x101)2 = multiply the Factor by 100 (1x102)Least Significant DigitSet the graduation.Settings of 0.5, 1, 2, 5, 10, 100 are available.The Custom Unit’s name can be customized up to 3 characters.Note: Custom Unit is locked at Off position when the Security Switch is set to the locked position.www. .com information@itm.com1.800.561.8187