Field CommentReanalysis Enables/disables the reanalysis feature, which allows you to adjustthe following ECG measurements:• P Duration• PR Interval• QRS Duration• QT IntervalThis is available only if Audit Trail is disabled and one of the follow-ing options is activated: ME12, MEHR, MI12, or MIHR.For more information on activating options, see 12.13 Options Set-up on page 151.QTC Calculation Determines which formula is used to correct QT calculations. Availa-ble options are:• BazettQTc = QT √HR\60Bazett is available only if the MEHR or MIHR option is activated.• FraminghamQTc = QT + 154 (1 — 60/HR)Framingham is available only if the ME12 or MI12 option is acti-vated.• FridericiaQTc = QT3 √HR/60Fridericia is available only if the ME12 or MI12 option is activated.NOTEIn all formulas, HR = Heart Rate.Screening Criteria Enables/disables the inclusion of the screen criteria.This setting is available only if the MI12 option is activated.If this filed is enabled, Suppress reason statements field is availa-ble.Suppress normal statement Enable/disables the inclusion of the normal statement.This setting is available only if the MI12 option is activated.Suppress abnormal/borderline Enable/disables the inclusion of the abnormal/borderline state-ments.This setting is available only if the MI12 option is activated.Suppress all statements Enable/disables the inclusion of all statements.This setting is available only if the MI12 or MIHR option is activated.Suppress reason statements Enable/disables the inclusion of reason statements.This setting is available if the MI12 option is activated.Continues on the next pageOperator Manual 12.2 Resting ECG Setup2053535-002 Revision T MAC 2000 ECG Analysis System 112/198