C Technical SpecificationsC.1 System SpecificationsTable C-1 Instrument TypeMicroprocessor augmented automatic electrocardiograph; 10-leadwire, 12 lead simultaneous acquisition with pro-grammable lead configuration.Table C-2 ProcessingItem SpecificationsECG Interpretation Marquette 12SL ECG Analysis Program for Adults and PediatricsComputerized Measurements 12-lead analysisECG Analysis Frequency 500 or 1000 samples/second/channelDigital Sampling Rate 16000 samples/second/channel for normal data acquisitionPace Sampling Rate 75K samples/second/channelECG On-screen Preview On-screen preview of acquired 10 second ECG waveformAcquisition Mode Provides 10 seconds of instantaneous ECG acquisitionDynamic Range AC Differential ± 5 mV, DC offset ±300 mVResolution 4.88 μV +/-1% per LSB @ 500SPSFrequency Range 0.04 to 150 HzLow Cutoff Frequency 0.04 Hz (ADS Off)0.56 Hz (ADS On)High Cutoff Frequency Configurable at 20 Hz, 40 Hz, 100 Hz or 150 HzCommon Mode Rejection >135 dB (with 50/60 HZ filter ON)Input Impedance >10MΩ @ 10 HzPatient Leakage <10 μA (Normal Condition), <50 μA (Single Fault Condition)Lead Detection All disconnected lead detection except RL & RAHeart Rate Meter 30 to 300 BPMStart-up Time Less than 30 secondsTable C-3 Patient InformationItem SpecificationsSupported patient Information Patient ID, Secondary Patient ID, Visit ID, Last name, First name, Height,Weight, Gender, Race, Pacemaker Patient, Systolic BP, Diastolic BP, Location#,Room, Order Number, Phone Number, Medication, Ordering physician, Refer-ring physician, Attending physician, Technician, Test indication.C2053535-002 Revision T MAC 2000 ECG Analysis System 186/198