Field CommentACI-TIPI Enables/disables the inclusion of the ACI-TIPI (Acute Cardiac Ische-mia Time Insensitive Predictive Instrument) statement and enablesthe Chest Pain field on the Patient Information window.To include ACI-TIPI statements, the following conditions must bemet:• MI12 or ME12 system option is activated• TIPI system option is activated• ACI-TIPI is enabled• 10s ECG Report Format is enabled• Print Interpretation is enabled• Patient data includes: gender, date of birth, and chest pain indi-cation• Patient cannot be a pediatric patient (15 years or younger) ascalculated form the date of birthSample Rate Determines the report frequency. Options are 500 Hz or 1000 Hz.1000 HZ is supported only for XML output.Field DescriptionLead Sequence Determines the lead sequence to use. Values are:• Standard• Cabrera• NEHB• SEQ4SEQ4 allows you to configure a custom 12-lead sequence using thefollowing fields. If either 12SL option (ME12 or MI12) is activated,you must select leads I (-I), II (-II), V1, V2, V3, V4, V5, and V6 for acorrect 12SL analysis.Sequence Name Set the display name for a custom lead sequence. Available only ifSEQ4 is selected for the Lead Sequence.Continues on the next pageOperator Manual 12.2 Resting ECG Setup2053535-002 Revision T MAC 2000 ECG Analysis System 113/198