Option DescriptionPrint Summary Prints a combined report that includes both the Table and Episode formats.Print Table Prints a breakdown of the recording in tabular format. The report includes:• the analysis duration in minutes and seconds• the artifact duration in minutes and seconds• a code for each event type recorded• the number of each event type recordedFor a description of the possible event codes, refer to 8.3 Arrhythmia Co-des on page 83.Print Episodes 'Prints a standard waveform report of the recorded events. The signal from allrecorded leads is printed and each event is marked with the corresponding ar-rhythmia code.For a description of the possible event codes, see 8.3 Arrhythmia Co-des on page 83.Main Menu Exits the Arrhythmia function and returns to the Main Menu.More Toggles between the arrhythmia recording options and the arrhythmia printingoptions.8.3 Arrhythmia CodesThe following table identifies the codes used on the Arrhythmia reports and the events they represent.Code Arrhythmia EventA ArtifactASYSTO Asystole, limit value 3sCPLT Ventricular couplet (2 PVCs)ESC Ventricular escape beatL Learn phasePAU1 Pause of 1 missed beatPAU2 Pause of 2 missed beatsPCAP Pacemaker capturePERR Pacemaker errorPSVC Premature supraventricular contractionPVC Premature ventricular contractionQRSL Learned QRS complexRUN Ventricular run (3 PVCs)VBIG Ventricular bigeminyVFIB Ventricular fibrillation/flutterVTACH Ventricular tachycardia (>3 PVCs)Operator Manual 8.3 Arrhythmia Codes2053535-002 Revision T MAC 2000 ECG Analysis System 83/198