Table 8-1 Arrhythmia Options - First Row (Table continued)Option Description20 HzNOTEThe initial measure-ment displayed is setin System Configura-tion > ArrhythmiaSetup.Eliminates noise in the waveform by restricting which frequencies are included.Frequencies are measured in Hertz (Hz) and include the following options:• 20 Hz• 40 Hz• 100 Hz• 150 HzSelecting a frequency eliminates signals that exceed that frequency. The smallerthe frequency selected, the more signal is filtered out. For example, a filter of 40Hz displays only signals at 40 Hz or less; signals greater than 40 Hz are ignored.Changing the measurement here does not change the measurement set in Sys-tem Configuration.More Toggles between the first row of options (previous) and the second row of options(following).Table 8-2 Arrhythmia Options - Second RowOption DescriptionPace Enhance Improves the readability of pacemaker ECGs. Options are On and Off.Patient Data Opens the Patient Data Entry window.This tab is available only if you did not complete the Portent Data Entry windowearlier.Main Menu Exits the Arrhythmia function and returns to the Main Menu.More Toggles between the first row of options and the second row of options8.2 PrintingYou can manually generate an arrhythmia printout in a table format, an episode format, or a summaryformat.8.2.1 Printing an Arrhythmia ReportUse the following procedure to print an Arrhythmia report.1. Select the type of Arrhythmia report you want to print and press the appropriate function key.• To print the summary report, press Print Summary.• To print the table report, press Print Table.• To print the arrhythmia episodes, press Print Episodes.Refer to 8.2.2 Arrhythmia Printing Options on page 82 for details.2. Review the report as necessary.For more information, refer to 8.3 Arrhythmia Codes on page 83.8.2.2 Arrhythmia Printing OptionsWhen printing an arrhythmia report, you have the following options:Operator Manual 8.2 Printing2053535-002 Revision T MAC 2000 ECG Analysis System 82/198