• Verify the ACI-TIPI option is activated.For information on activating the ACI-TIPI option, refer to 12.13 Options Setup on page 151.• Verify ACI-TIPI is enabled on the ECG.For information, refer to 12.2 Resting ECG Setup on page 108.• Verify the information ACI-TIPI requires was entered.The ACI-TIPI statement prints only if the patient’s gender, date of birth, and chest pain indication areincluded in the patient information.• Verify the patient is 16 years old or older.The ACI-TIPI statement does not print for pediatric patients.• Verify the original ECG was acquired in an electrocardiograph with the ACI-TIPI option.If you attempt to print an ECG that was imported from an external device, the cart does notgenerate an ACI-TIPI statement; it prints only if the statement was saved as part of the ECG.15.7 System ErrorsThe following table identifies some potential errors that may occur while you are operating the system,the possible causes, and a recommended course of action to resolve the error.If performing the recommended actions does not resolve the problem, contact authorized GEHealthcare service personnel.Problem Cause SolutionThis icon is displayed andthe battery LED is flash-ing.System is operating from the bat-tery and the battery charge is low.Connect the system to an AC outlet to charge thebattery.This icon is displayed andthe battery LED is not lit.System is operating from AC pow-er and battery is not installed.Install a battery.The system does not pow-er up while operating frombattery power.Battery is fully discharged. Connect the system to an AC outlet to charge thebatteryThe system powers downwhile operating from bat-tery power.Battery is fully discharged Connect the system to an AC outlet to charge thebatteryContinues on the next pageOperator Manual 15.7 System Errors2053535-002 Revision T MAC 2000 ECG Analysis System 176/198