7.3 Full DisclosureFull Disclosure allows you to acquire and save up to five minutes of ECG waveforms to print at a latertime or view on a computer. You can manually cancel or finish acquisition at any time before the fiveminutes are concluded. Full Disclosure is optional and can be configured during system setup.This section outlines the procedure for generating a Full Disclosure ECG report and describes theavailable setup, waveform, and output options.7.3.1 Generating a Full Disclosure ECG Report1. Configure the Full Disclosure setting in 12.2 Resting ECG Setup on page 108.2. Prepare the patient as described in 4 Preparing the Patient on page 53.3. From the Main Menu, press Resting ECG > More > Full Disclosure to open the Full Disclosurewindow.The system automatically starts acquiring the full disclosure data, and the Acquiring messagedisplays on the screen during the acquisition.The system automatically completes the test after the pre-configured time, and you can continuewith step 6.Otherwise, go to step 4 or step 5.4. Press Finish to complete acquiring the ECG. The following Full Disclosure Report windowdisplays.Continue with step 6.5. Press Cancel to stop acquiring the ECG, and perform one of the following:• Press Yes to stop acquiring the ECG data and return to the Resting ECG window.• Press No to remain acquiring the ECG data, and continue with step 6.6. In the Full Disclosure Report window, you can perform the following operations:Full Disclosure Report Window Field ActionExport to SD Card Press to export the ECG report to the SD card.Continues on the next pageOperator Manual 7.3 Full Disclosure2053535-002 Revision T MAC 2000 ECG Analysis System 75/198