- 17 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setIntended Use5The FLS 106 scanner is a positioning system for probes from Langer EMV-Technik GmbH and isused exclusively to test and/or carry out measurements on printed circuit boards (PCBs) andintegrated circuits (ICs).The system supports the following probe types from Langer EMV-Technik GmbH:- ICR near-field micro-probes- Passive near-field probesThe system allows the measurement of fields coupled out by the device under test (DUT).The probes can be moved across the DUT surface on the X-, Y- and Z-axis.In addition, the FLS 106 IC allows the probe's rotation around the Z-axis. The probe tip positioncan be checked visually with the DM-CAM digital microscope camera at any time.The FLS 106 is controlled with the ChipScan-Scanner software via a computer.The system is used for applications such as:- Surface scans in compliance with standard DIN IEC/TS 61967-3- Volume scans- Pin scansThe FLS 106 scanner may only be used in accordance with the prerequisites described insection 6.1.The user must comply with all information and instructions given in this user manual.The FLS 106 scanner may only be used in an environment with a temperature of10 to 30 degree Celsius and a humidity of 20 to 85 percent without condensation.Keep the scanner free of foreign objects, dirt and liquids to prevent adverse effects onmeasurements and damage to the scanner.Staff Requirements5.1Only persons who are trained and qualified in the field of electromagnetic compatibility (EMC) mayoperate the FLS 106 scanner.Persons whose responses are impaired by alcohol, medicines, drugs or similar substances mustnot be allowed to operate the FLS 106 scanner.