- 34 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setSH 01 Probe Holder10.8The SH 01 probe holder is designed to hold a near-field probe. The SH 01 comprises a mountingbracket, the magnetic holder and the probe sleeve.The probe sleeve is held magnetically and can be quickly loosened to easily install a probe, forexample.The probe sleeve has a bore on the rear so that it can be returned into its correct position on themagnetic holder by simply positioning the bore directly above the magnetic holder's pin.Figure 33: SH 01 comprising the mounting bracket, magnetic holder and probe sleeveIn addition, the SH 01 has a collision protection function: The probe moves up with the magnetic holder if the probe touches an obstacle when movingdown. The scanner can recognize this displacement and stop the movement. The probe sleeve with the probe is released from the magnetic holder and falls off if the probehits an obstacle when moving sideways. This normally prevents any damage to the DUT.Please read section 18.3 on how to set up the collision protection when the probe is moving down(collision control).Figure 34: SH 01's magnetic holder and probe sleeve with SSMB connector and clamping screw to hold the near-fieldprobe