- 66 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setFigure 85: Connecting the SMA-SMA cable Figure 86: Connecting the power supply cableConnect the bias tee’s power supply cable to the input of the bias tee marked 12 V / DC and to theconnector on the Z-axis tower marked 12 V DC (HR10 6-pin connector, blue) (Figure 86).Collision Control of the ICR Probe17.5The collision control is a protective function in connection with an ICR probe to limit potentialdamage that may be caused if the ICR probe accidentally touches the device under test.The ICR probe may not touch the device under test during operation. The collision control may notbe used to position the ICR probe.The collision control has to be activated in the ChipScan-Scanner software by ticking theDepth Test checkbox. This is only possible if the SSMB-SSMB depth-test cable is connected to theSSMB connector of the ICR probe and the SSMB connector on the rotary unit (Figure 73,Figure 80) and the ICR probe's power supply has been established via the BT 706 bias tee.With the collision control function activated, the rotary unit's downward movement is stopped if theICR probe tip touches the surface of a DUT, for example.