- 20 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setSafety Devices6Prerequisites for Safe Operation6.1The FLS 106 scanner may only be operated if either- a protective enclosure or- a spatial separation in connection with the external NA 5 EMERGENCY STOP switch isused.The FLS 106 scanner is stopped immediately if- the EMERGENCY STOP switch on the device is actuated,- the external EMERGENCY STOP switch is actuated,- the protective enclosure is opened.After the EMERGENCY STOP switch on the scanner has been pressed, it has to be released byturning it clockwise before the scanner can be switched back on.After the external NA 5 EMERGENCY STOP switch has been pressed, it has to be released bypulling before the scanner can be switched back on.The scanner then has to be recalibrated with the software.Operation with the Protective Enclosure6.1.1The FLS 106 scanner is operated inside the SUH 106 protective enclosure.Operation with a Spatial Separation6.1.2The user provides a spatial separation at an appropriate distance to prevent persons frominterfering with the scanner's operation. Measures such as cordoning off or fencing off the site aresuitable precautions. The external EMERGENCY STOP switch has to be connected to the scannerand placed outside the isolated space so that it is accessible at any time.Switching the Scanner on in Normal Operation and after a Power6.2FailureThe FLS 106 scanner can only be activated or operated via the software after passing a self-testing routine.EMERGENCY STOP Switch6.3The emergency stop function is triggered by pressing the EMERGENCY STOP switch.The FLS 106 scanner is stopped immediately.After the EMERGENCY STOP switch has been pressed, it has to be released by turning itclockwise before the scanner can be switched back on.Overcurrent Protective Device6.4Line-side fuses protect the scanner in the event of a fault. The power supply unit is resistant toshort circuits.