- 58 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setPlace the clamping pieces above the T-nuts and secure loosely with the hexagon socketscrew M6.Figure 64: Fastening the UH DUT with the clamping pieces on the T-slot tableAdjust the height and angle with the grub screws as required. Fix the clamping pieces with thehexagon socket screw M6.After installing the UH DUT universal holder, check that this cannot slip. If necessary, furthertighten the screws M8 with care.Fastening the Circuit Board16.2.2Place the circuit board in the middle of the UH DUT.Use either claw 01 or claw 02 depending on the height of the board. Claw 01 is designed to hold acircuit board with a height of less than 8 mm. Claw 02 is designed to hold a circuit board with aheight of less than 14 mm.The claws can be adjusted continuously and have to be fixed with the screws M3 (length of 12 mm,hexagon socket of 2.5 mm) in the holes of the UH DUT.The circuit board should be fastened with two claws each on two opposite sides (Figure 65).Figure 65: Fastening example with claw 02The claws have to be fastened so that the circuit board cannot be moved.The remaining claws can also be used for this purpose.