- 42 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setPutting the FLS 106 Scanner into Service for the First13TimeRead Chapter 12 before putting the system into service for the first time.Standard procedure:a) Install and connect the digital microscope camera:- refer to section 14.3.2 for FLS 106 IC or- refer to section 14.4 for FLS 106 PCBb) Check that the cable of the rotary unit is connected correctly (connect if necessary)(Figure 40).c) Connect the mains cable (Figure 49).d) Ensure the system's safe operation:- set up the spatial separation and connect the external NA 5 EMERGENCY STOP switch(section 14.1.2) or- set up and connect the SUH 106 safety enclosure (section 14.1.1)e) Check the EMERGENCY STOP switch and release if necessary (section 12.8).f) Connect the measuring device to the SMA connector on the control panel (please refer tothe list of measuring devices supported by the CS-Scanner software) (section 14.7).g) Connect the scanner to the computer via a USB type A/B cable (section 14.6).h) Close the SUH 106 safety enclosure if this option has been chosen.i) Switch on and boot the computer.j) Switch on the scanner and install the driver on the computer (section 15.2).k) Install the ChipScan-Scanner software on the computer (section 15.3).l) Start the ChipScan-Scanner software and detect the devices that are connected(section 15.4).m) Calibrate and test the scanner via the ChipScan-Scanner software (section 15.4 item 4).