- 38 / 83 -LANGEREMV-TechnikDE-01728 Bannewitzmail@langer-emv.dewww.langer-emv.de FLS 106 IC set / FLS 106 PCB setThe FLS 106 scanner may only be set up and used in closed rooms under appropriate lightconditions.Make sure that the free working area around the scanner is large enough for the scanner tobe handled properly.Take the space requirements for cordoning off the area into account if you intend to operatethe system with a spatial separation.The scanner should be installed on a raised surface such as a table, workbench, etc. Checkthe worktop's size and load-bearing capacity before installing the system. The surface mustbe level, firm and clean.If you intend to operate the system with the SUH 106 safety enclosure, ensure that theworktop is large enough and its load-bearing capacity is not exceeded (total weightincluding the SUH 106 approx. 125 kg).Secure the scanner and all individual components against falling down.Damage to property due to incorrect tightening of screws!All components for which a special tightening torque is defined have already been installedand tightened. Do not tighten these screws afterwards!The screws of all attachments that have to be installed later should only be fastened hand-tight. No torque wrench is needed for this purpose.Setting up the FLS 10612.3Place the FLS 106 scanner at its place of operation. Follow the instruction in Section 12.2.Acclimatization of the FLS 106 Scanner12.4Give the scanner enough time to acclimatize before putting it into service if it is brought from a coldinto a warm environment. Condensation may settle down on the scanner's surface. Thecondensation should be promptly removed with dry, clean cloths. Acclimatization may take up toseveral hours depending on the temperature difference. We recommend that further steps only betaken after the scanner has adjusted to the working area's ambient temperature.!Caution!