23Magtrol Model DSP7000 Dynamometer Controller C h a p t e r 3 – Installation/CongurationSETUP3.2.3 H ySTereSiS DynaMoMeTer wiTH TranSDuCer SeTup3.2.3.1 Hardware ConnectionMotorUnderTestTM Torque TransducerAC MainsHysteresis Dynamometer (HD)DSP7002DYNAMOMETERCONTROLLER No ConnectionUSBPCM-TESTFigure 3–10 Hysteresis Dynamometer with Torque Transducer Setup3.2.3.2 Software Configuration1. Turn on the DSP7000. Set up TSC1 as described in section 3.2.2 - Hysteresis DynamometerSetup. Press TSC to switch to TSC2 setup and proceed to the dynamometer configurationmenu. See Section 3.2.1 – Dynamometer Configuration Menu.2. Press PoWER UNITS until TM/TF is reached for TSC2.3. Press ToRqUE UNITS until hb is reached.4. Press MAx SPEED button and use and buttons and Decrease/Increase dial to setdesired scale factor for TSC2.5. Press SCALE I to add a filter if desired.6. Press ShIFT. The display should appear as follows:POWERBRAKE STATUS SET POINT SET POINT P I DTORQUE SPEED USER DISPLAYTSC2 MAX TORQUE NOMINAL SPEEDX.XXX XXXX.XFigure 3–11 Torque Transducer Setup Menu7. Press the ToRqUE UNITS button and use and buttons and Decrease/Increase dial toset desired max torque. Press the SCALE P button and use and buttons and Decrease/Increase dial to set desired nominal speed.8. Press ShIFT 3 times to return to the main menu.