26Magtrol Model DSP7000 Dynamometer ControllerChapter3 – Installation/CongurationSETUP4.a If using an Eddy Current Dynamometer, press ShIFT. The display should appear as follows:POWERBRAKE STATUS SET POINT SET POINT P I DTORQUE SPEED USER DISPLAYTSC1 MAX TORQUE NOMINAL SPEEDXXXX XXXX.XFigure 3–16 TSC1 Eddy-Current Setup MenuPress the ToRqUE UNITS button and use and buttons and Decrease/Increase dial toset desired max torque. Press the SCALE P button and use and buttons and Decrease/Increase dial to set desired nominal speed. Press ShIFT 3 times to complete the initial setupand return to the main menu.4.b If using a Powder brake Dynamometer, press ShIFT. The display should appear as follows:POWERBRAKE STATUS SET POINT SET POINT P I DTORQUE SPEED USER DISPLAYTSC1 MAX TORQUEXXXXFigure 3-17 TSC1 Powder Brake Setup MenuPress the ToRqUE UNITS button and use and buttons and Decrease/Increase dial toset desired max torque. Press ShIFT 3 times to complete the initial setup and return to themain menu.