Azbil Corporation Maintenance and TroubleshootingModel MGG10C/14C - MagneW FLEX+/PLUS+ Converter 5-5Contact input/output loop checkBy switching contact input terminal ON/OFF, contact input terminal status can bechecked through the LCD.By switching contact output terminal ON/OFF, contact output loop can be checked.Display varies according to the selection of contact input and output.Default settingDisplay the existing statusSetting range“CLOSE”, “OPEN”Step Screen Procedure(English) (Japanese)1 > ST. IN OPENST. OUT CLOSE>4 - 20 000.0 %Enter MAINTENANCE MODE(see page 4-24). Use the orkey to cycle through screenuntil “OUTPUT CHECK”screen appears. Touch thekey once to move the cursorto “OFF”. Touch the key toswitch the OFF screen to ONscreen. Touch the key twiceto move to “OUTPUT CHECKST. IN/OUT” screen.2 > ST. IN OPENST. OUT CLOSE>4 - 20 000.0 %Touch the key to move thecursor at the status (OPEN/CLOSE).3 > ST. IN CLOSEST. OUT OPEN>4 - 20 100.0 %Use the key to select thedesired output status (OPEN/CLOSE).4 > ST. IN CLOSEST. OUT OPEN>4 - 20 100.0 %For the contact input, the displayindicates current contact inputstatus (OPEN/CLOSE).