Installation Azbil Corporation2-6 Model MGG10C/14C - MagneW FLEX+/PLUS+ ConverterIntegral wiring - 1 (1-contact input and 1- contact output)To wire a remote system, the following cables are required:• Analog output cable - see page 2-16• Pulse output cable - see page 2-17• Contact input/output cable - see page 2-17The following pages provide information on selecting the correct cables and wiringthe system. A diagram of the terminal block for a remote system is shown below.1-contact input and 1-contact outputACPOWERE STATUS IN STATUS OUT+ - + -+ - + -POWER ACL N I. OUTP. OUTE STATUS IN STATUS OUT+ - + -+ - + -POWER ACL N I. OUTP. OUTSB SA C ABX YSB SA C ABX YFigure 2-4 Integral converter terminal blockTable 2-1 Remote converter terminal descriptions (1-contact input & 1-contact output)Symbol DescriptionI. OUT + Analog output–P. OUT + Pulse output–STATUS IN + Contact input–STATUS OUT + Contact output–E Not usedGrounding (grounding resistance must be <100W)~ Note In case of DC24V and DC110V power supply, the symbol of the “L” and “N” of thepower supply become “+” and “–”.