Using the display panel Azbil Corporation4-42 Model MGG10C/14C - MagneW FLEX+/PLUS+ ConverterSetting hysteresisThis is used to set the hysteresis as a range function to be used at range switching.Use for normal direction automatic dual range or the normal/reverse directionautomatic range.Default setting0%Setting range0-20%Step Screen Procedure(English) (Japanese)1 HYSTERESIS05 % 05 %Open the hysteresis setup screenby following the steps to enterthe engineering mode. (See page4-11)2 HYSTERESIS05 % 05 %Touch the key once.3 HYSTERESIS10 % 10 %Use the and keys to inputthe desired hysteresis value.In this example, here thehysteresis is changed from 5%to 10%.4 HYSTERESIS10 % 10 %Touch the key to move thecursor to the “#”.