Maintenance and Troubleshooting Azbil Corporation5-6 Model MGG10C/14C - MagneW FLEX+/PLUS+ ConverterStep Screen Procedure(English) (Japanese)5 > ST. IN CLOSEST. OUT OPEN> 4 - 20 100.0 %Touch the key to move thecursor to “>”.6Use the or key to moveother screen. Then the contactinput and output are back to thecurrent status (OPEN/CLOSE).<2-contact inputs>Step Screen Procedure(English) (Japanese)1 > ST. IN1 CLOSEST. IN2 CLOSE> 1 CLOSE2 CLOSEEnter MAINTENANCE MODE(see page 4-24). Use the orkey to cycle through screenuntil “OUTPUT CHECK”screen appears.Touch the key once to movethe cursor to “OFF”.Touch the key to switch theOFF screen to ON screen.Touch the key twice to moveto the “OUTPUT CHECK ST.IN” screen.2 > ST. IN1 CLOSEST. IN2 OPEN> 1 CLOSE2 OPENThe display indicates the currentcontact input status (OPEN/CLOSE).