Using the display panel Azbil Corporation4-18 Model MGG10C/14C - MagneW FLEX+/PLUS+ ConverterDetector dataThis function is used to select the constant, model and diameter of the detector beingused with this converter. In this screen, three values can be changed: the detectorconstant (EX), detector type (MGG, KID, NNK, NNM) and the detector diameter(DIA)Default settingIf only converter is purchased, the default values are EX 300.0, MGG, DIA050.0.CAUTIONIf the converter and detector are purchased together, the converter contains thedetector data that was set during actual flow calibration. DO NOT change this dataor the flowmeter output will be incorrect.Step Screen Procedure(English) (Japanese)1 EX 300.0MGG DIA 050.0Enter BASIC SETUP MODE(see page 4-11).Use the or key to cyclethrough the screens until thedetector DATA (EX/DIA)screen appears.2 EX 320.0MGG DIA 050.0Touch the key until thecursor is at the detector Constant(EX) to be changed.Touch the or key tochange the value.3 EX 320.0KID DIA 050.0Touch the key until thecursor is at the detector type.Use the or key to changethe value.4 EX 320.0KID DIA 100.0Use the key to move thecursor to the detector Diametervalue.The converter works with thefollowing detector Diameters (inmillimeters):s 2.5 s 5 s 10 s 15s 25 s 40 s 50 s 65s 80 s 100 s 125 s 150s 200 s 250 s 300 s 350s 400 s 450 s 500 s 600s 700 s 800 s 900 s 1000s 1100