Azbil Corporation InstallationModel MGG10C/14C - MagneW FLEX+/PLUS+ Converter 2-11Remote wiring - 3 (2-contact output)To wire a remote system, the following cables are required:• Signal cable - see page 2-12• Excitation cable - see page 2-12• Analog output cable - see page 2-16• Pulse output cable - see page 2-17• Contact input/output cable - see page 2-17The following pages provide information on selecting the correct cables and wiringthe system. A diagram of the terminal block for a remote system is shown below.2-contact outputACPOWERE STATUS OUT 2 STATUS OUT 1+ - + -+ - + -X YPOWER ACL NSB SA C ABI. OUTP. OUTE STATUS OUT 2 STATUS OUT 1+ - + -+ - + -X YPOWER ACL NSB SA C ABI. OUTP. OUTFigure 2-9 Remote converter terminal blockTable 2-6 Remote converter terminal descriptions (2-contact output)Symbol DescriptionAFlow rate signal inputBCSASBI. OUT + Analog output–P. OUT + Pulse output–STATUS OUT 1 + Contact output 1–STATUS OUT 2 + Contact output 2–X ExcitationYE Not usedGrounding (grounding resistance must be <100 W)~ Note In case of DC24V and DC110V power supply, the symbol of the “L” and “N” of thepower supply become “+” and “–”.