Azbil Corporation InstallationModel MGG10C/14C - MagneW FLEX+/PLUS+ Converter 2-9Remote wiring - 1 (1-contact input and 1- contact output)To wire a remote system, the following cables are required:• Signal cable - see page 2-12• Excitation cable - see page 2-12• Analog output cable - see page 2-16• Pulse output cable - see page 2-17• Contact input/output cable - see page 2-17The following pages provide information on selecting the correct cables and wiringthe system. A diagram of the terminal block for a remote system is shown below.1-contact input and 1-contact outputACPOWERE STATUS IN STATUS OUT+ - + -+ - + -X YPOWER ACL NSB SA C ABI. OUTP. OUTE STATUS IN STATUS OUT+ - + -+ - + -X YPOWER ACL NSB SA C ABI. OUTP. OUTFigure 2-7 Remote converter terminal blockTable 2-4 Remote converter terminal descriptions (1-contact input & 1-contact output)Symbol DescriptionAFlow rate signal inputBCSASBI. OUT + Analog output–P. OUT + Pulse output–STATUS IN + Contact input–STATUS OUT + Contact output–X ExcitationYE Not usedGrounding (grounding resistance must be <100W)~ Note In case of DC24V and DC110V power supply, the symbol of the “L” and “N” of thepower supply become “+” and “–”.