Model MGG10C/14C - MagneW FLEX+/PLUS+ Converter 3-1Chapter 3: OperationThis chapter describes the procedure for starting and shutting down the flowmeter andusing the display panel and the infrared touch sensor keys.Start-upTo start operation of the flowmeter:1. Confirm that the detector is correctly installed on the pipe.2. Confirm that the wiring between the converter and the detector has been properlycompleted according to the installation instructions specified in this manual. Incase of using a HART/SFC communication function, verify an appropriate wiringaccording to the wiring in this users manual.3. Begin and then stop fluid flow through the detector so that fluid is present in thedetector in a static condition.4. Confirm that there is no leakage at the flanges of the detector.5. Apply power to the converter.6. The following display appears seven seconds after the power is ON.:WPO 0000000020Figure 3-1 Start-up display7. Zero the flowmeter using the procedure on page 4-14.The flowmeter is now on and operational.Shut downCAUTIONSwitch the control equipment to manual control before terminating flowmeteroperation and shutting off the output to the control equipment. This actionprevents the power shut-off from directly affecting the control equipment andcausing the valve positioner to malfunction.To stop operation of the flowmeter:1. Switch the control equipment connected to the flowmeter to manual control.2. Turn off power to the converter.