Azbil Corporation Using the display panelModel MGG10C/14C - MagneW FLEX+/PLUS+ Converter 4-37[Contact output function]This function can be set when 1- or 2-contact output has been selected in theadditional specifications.X: Not activated1: Alarm contact outputAn alarm is output when any of the following items becomes abnormal.The abnormal item can be checked on the display of the instrument.Also, external confirmation is available using the SFC.j Self-diagnostic• Coil disconnection• ROM error• RAM error• NVM error• ADC errorTable 4-2 Output selectionModeselectionFail-safe direction high(HIGH)Hold(HOLD)Fail-safe direction low(LOW)Analogoutput4 to 20 mAFail-sage high (HIGH)without SFCcommunication:24 mA DCwith SFC communication:23.8 mA DCHold(HOLD)Fail-safe direction low(LOW)Without SFC communication:0.8 mA DCWith SFC communication:2.96 mA DCPulseoutputHold(HOLD)Fail-safe direction low(LOW)No outputContactoutput Abnormal status (Open/closed can be freely selected.)CAUTIONIf the power supply is turned OFF with the “Fail-safe direction high” setting,the 4-20 mA output will emit a Fail-safe direction high output once. Pay closeattention when turning the power supply OFF.