CHAPTER 1 - SYSTEM OPERATIONLSM 880 Center Screen Area / Image Containers - Display and Image Analysis ZEISS10/2014 V_01 000000-2071-464 5236.19.6 Model FunctionalitiesActivate Model under the RICS view. The Model options will be displayed in the View control window(Fig. 736).Fig. 736 Fit control windowWith the Model pull down menu you can select asaved pre-defined model from the list. You canrename a created or changed model by typing in anew name in the display box. If the Ok button ispressed, the model will be saved under the newname (Fig. 737). If the model name was notchanged, you will be prompted to acknowledge tooverwrite the old existing one in the Save ModelWarning menu (Fig. 738). Press Yes to overwrite,No to keep the old model.Type in a name in the Model name display boxand press Ok, if you want to save the model, orpress Cancel without saving. You can delete amodel by pressing the Cross button. You will beprompted to acknowledge the action within thepopped up Delete Model menu (Fig. 739). PressYes to delete, No if the model should be kept.Below the Model pull down box the availableterms to assemble a model are listed. If you wantto take a term into account, check thecorresponding box. Four terms are available:Scanning Gs, Offset d, Amplitude A andTranslation Gd. The assembled model will bedisplayed in the spatial correlation G(ξ,ψ) displayarea.Fig. 737 Model Save menuFig. 738 Save Model Warning menuFig. 739 Delete Model menu