264RS-232C Communication Chapter 11Transmitting Product Types and Environment Data from a Compact Flash MemoryCard to the PV310 (%F)This command transmits the settings and configuration data saved on the Compact Flash memory card to PV310using the signals from an external device such as PLC. This function will be helpful if more than 64 product typesare needed. The product types and configuration data are stored in the folders (Program 01 to Program 99) underPanasonic MEW Vision¥PV310¥Data.Communication direction Command / DescriptionTransmission fromexternal device % F ? ? CR??: 01 to 99 (Specify the number of the folders where product types and configurationdata are stored in 2-digit form.)Response: Normal(at the end of transmitting) % F CRResponse: Error 1 % U CRCause:Command in the unspecified form or not provided was received.Examples:- %FCR (any other checker number than 01 - 99 was specified. (1 or 3-digitnumbers were specified.- %B01CRResponse: Error 2• ERROR signal = “ON”• Error Code: E0173 isdisplayed on thescreen% Z CRCauses:• Memory card is not inserted into the PV310.• The specified folder does not exist in the path.(Panasonic MEW Vision¥PV310¥Data).• Configuration data file "TYPEENV AXD" does not exist in the specified folder.