INTRODUCTION TO LASER SCANNING MICROSCOPYLSM 510 FunctionsB 40-051 e 07/98 5-251 - Wrap No normalization - the grey values are displayed modulo 256 (4096). If the resultis less than 0, the value 256 (4096) is added to it.2 - Clip Negative values are set to 0.3 - Normalize The resulting grey value range is scaled to the range 0...255 (0...4095).4 - Shift/Clip 128 (2048) is added to the difference, then negative values are set to 0. Valuesgreater than 255 (4095) are set to 255 (4095).Parameters:Input 1 First input image sequenceInput 2 Second input image sequenceOutput Output image sequenceMode 1 - Wrap2 - Clip3 - Normalize4 - Shift/Clip