OPERATIONLSM 510 Image Acquisition (Acquire)B 40-051 e 07/98 4-31With Transmitted Light activated , the halogen lamp is automatically occluded in the laser scanningmode.Please bear in mind that the light intensity does not automatically correspond to 0 % when LightRemote is deactivated. The microscope setting (light intensity) of the last session, which was notremote-controlled, is restored on exit of the program. In the Objectives list box, select the required objective as follows: Open the list box. Click on the objective you want to select. The selected objective will automatically move into the beam path.+ Note that the nosepiece of the Axiovert 100 M microscope must not be moved manually. Tobring another objective into the beam path, move the objective nosepiece only software-controlled.Fig. 4-33