LSM 510 - SETUP REQUIREMENTSPower Requirements / Physical Dimensions LSM 5102-6 B 40-051 e 07/982.2.1 Phase 1 (LSM)feeds the following units:Laser ModuleHeNe 2xvia Power 1 (5-socket adapter)Computer + monitorMicroscopeMCU 28Scanning Modulevia Power 2:HAL lampHBO lamp2.2.2 Phase 2 (LSM, Power 3)feeds the following units:Ar laser 2 kWor ArKr laser 2 kW2.2.3 Separate connection:Ar laser (UV) 7 kW2.3 Physical DimensionsLength (cm) Width (cm) Height (cm) Weight (kg)Large system table 150 80 78 100Small system table 65 80 78 60Scanning Module 25 20 25 15Microscope 50 35 50 20Laser Module, VIS(ible light) 80 40 50 60Laser Module, UV 140 20 20 60Electronics box 50 30 30 10Power supply for Ar, ArKr 30 30 20 10Power supply for Ar (UV) 50 50 30 30Cooling unit for Ar (UV) 80 45 50 30Water hose for Ar (UV) 400Fiber optic cable, VIS(ible) 200Fiber optic cable, UV 200Cables 250SCSI cable 100