LSM 510 OPERATION IN EXPERT MODELSM 510 META Acquire Menu Carl Zeiss03/06 B 45-0021 e 4-115For Z Stacks over the time (4D) following offline functions will be enlarged:− Slice (Z slider and Time slider)− Gallery (Z, Time and Z + Time buttons)− 3D (slider for single time index)To select the Z or Time Slices, use the appropriate sliders which are displayed if the Slice button in theImage Display window has been activated.When you click on the Z + Time button, all individual frames will be displayed.Working in Online Fingerprinting mode when acquiring a Z Stack results in a similar imagewindow with the same display functionalities.4.5.6.8 Time Series of a Frame over a Z Stack over Lambda• Activate the META mode sheet in the configuration control and set the relevant parameters.• Set the relevant parameters for the Start Series, End Series and Time Delay panels.• Start the time series with a click on the Start T or Start B button.Fig. 4-95 Image Display window of a Time Series over Lambda