LSM 510 OPERATION IN EXPERT MODELSM 510 META Purpose Carl Zeiss03/06 B 45-0021 e 4-114.1.3 Convention for the Text in this ManualAll the originally used terms of the software interface, e.g.− names of windows,− panels,− input boxes,− list / selection boxes,− check boxes,− menu items,− names of buttons and− keyboard keys,are displayed in bold letters to allow easier identification.4.1.4 BackupSystem backup− A complete backup is contained on the enclosed backup CD-ROM.User files backupThe following user-generated files need to be included in a backup procedure (keep directory structure):− Image database files: *.mdb (but not system_configuration_*.mdb)− LSM Image files: *.lsm− Exported images: *.* (*.Tiff, *.LSM-Tiff, *.BMP, …)− Palette files: AIM \ Palette \ *.lut− Filter files: AIM \ Filter \ *.krn− Pinhole setting files: AIM \ PH*.pos− Log files: AIM \ *.logThe following files generated during the system integration should also be included in a backupprocedure:− Parameter file for pinhole setting: AIM \ *.set− Parameter file after pinhole adjustment: AIM \ *.adj− Scanner files: AIM \ bin \ *.bin− Microscope stand database: AIM \ database \ system_configuration_*.mdb4.1.5 Software OperationThe LSM 5 software can be operated using the mouse, the PC keyboard, or both.The operation of the mouse and the keyboard is identical to that of the Microsoft ® WINDOWSoperating system and is therefore not dealt with in detail in this manual.If required, see the Microsoft manual or online help for relevant information.