LSM 510 AND LSM 510 META - SETUP REQUIREMENTS LSM 510 DuoScanCarl Zeiss Vibrations / Microscopes LSM 510 META DuoScan2-16 B 45-0021 e 03/062.6 VibrationsVibrations under operation conditions(with system table)Shipping shock (LSM 510 box)5 μm pp at 5 Hz 3 g10 μm pp at 10 Hz10 μm pp at 20 Hz2.7 MicroscopesInverted Axiovert 200 M BP or SPUpright Axioskop 2 FS MOTUpright Axio Imager.Z1Upright Axio Imager.M1All ICS objectives from Carl Zeiss and their accessories can beaccommodated.Z motor DC servomotor, opto-electronically codedLeast Z interval: 50 nm (Axiovert 200 M BP or SP)100 nm (Axioskop 2 FS MOT )10 nm (Axio Imager.Z1)25 nm (Axio Imager.M1)Piezo Objective focus Piezo-driven single objective driveMax. travel 250 μm; resolution 10 nmIn the unlikely case of extreme fluctuations of the external power net orelectromagnetic radiation, the piezo crystal will vary and disturbance inthe image is visible. Note that this is not a defect and the piezo drive willnot be damaged.