LSM 510 NOTES ON DEVICE SAFETYLSM 510 META User Interface Carl Zeiss03/06 B 45-0021 e 1-25To ensure functioning of the system and laser safety the following connections have to be changed:1. The connection of the microscope to the safety interface of the system is located either on theadditional Safety-Box (Axioskop 2 FS MOT , Fig. 1-22/1) or on the rear side of the microscope (Axiovert200 M or Axio Imager.Z1; Fig. 1-23/1 and Fig. 1-24/1). This connection has to be unplugged fromthe microscope which is not in use and plugged into the microscope to be used following theexchange of the scan head.Fig. 1-22 Safety-Box of Axioskop 2 FS MOTwith main connection to safetyinterface (1)Fig. 1-23 Connection of Axio Imager.Z1to safety interface (1) andelectronics (2)1Fig. 1-24 Connection of Axiovert 200 M tosafety interface (1)Fig. 1-25 Four CAN connections (1) areavailable on the rear side of theelectronics module. The microscopein use has to be connected to oneof them